標題 Preservation of hydrogen dissolved in hemofiltration replacement fluid and hydrogen permeability of kemofiltwation membranes
執筆形態 共著
掲載誌名 Journal of Society for Technology of Blood purification
掲載区分国内
出版社・発行元 株式会社 春恒社
巻・号・頁 26(2),pp.151-160
総ページ数 10
著者・共著者 荻野稔、苗村潔、伊藤奈々、笠井亮佑、加納敬、上条史記、楠元直樹、秋本和哉、石川貴一、大脇杏実、西澤菜津子、田仲浩平、篠原一彦、横山憲二
概要 Introduction: Dialysis using hydrogen-dissolved solution has become an attractive method considering the
oxidative stress suppression effects of hydrogen. Here, we focused on the use of hydrogen-dissolved solution
for hemofiltration. We specifically evaluated the hydrogen-dissolved blood filtration replacement fluid
preservation method and the hydrogen permeability of the hemofiltration membranes. To the best of our
knowledge, this is the first report focusing on the hydrogen-dissolved hemofiltration replacement fluid.
Methods: To define the shelf life of the hydrogen in solution, we assessed the influence of dissolved hydrogen
concentration and the composition of the substitution solution stored at room temperature and under
refrigeration. Moreover, we measured the hydrogen concentration of the water hemofiltration replacement fluid
that passed through the hemofiltration membrane to assess the hydrogen permeability of the hemofiltration
membrane using the hemofiltration simulation circuit. We compared three kinds of hemofiltration membranes:
polymethyl methacrylate, sodium-acrylonitrile methallylsulfonate copolymer (AN 69 ST), and polysulfone
(PS) membrane.
Findings: The residual hydrogen concentration of the hydrogen-dissolved hemofiltration replacement fluid
stored for 4 weeks was 1% at room temperature( 23 ± 3℃) and 19% at refrigeration( 4℃) with respect to
the initial value. No changes were observed in the composition concentration of hydrogen-dissolved blood
filtration replacement fluid. Although all tested blood membrane filters demonstrated hydrogen permeability.