概要 | This paper presents the comparison ofsensitivity between the reflection and transmission coefficient by free space method for non-invasive blood glucose monitoring. In this study, aqueous glucose solutions with concentration values of 0, 1, 2, 3, and 4 w/v% have been tested. The calculation is done using formulas based on a three-layer model.
The calculation results are then comparedwith the simulation results. Both calculation and simulation show almost the same results. The measurement of reflection and transmission coefficient is also implemented by using horn antenna. The results show that transmission coefficient is more sensitive with respect to the change of glucose concentration. 副筆 |