標題 | Blood-Mimicking Fluid for Testing Ultrasonic Diagnostic Instrument(超音波診断装置を検査するための血液疑似流体) (査読有) |
執筆形態 | 共著 |
掲載誌名 | Japanese Jounal of Applied Physics.51(2012)07GF18-1-5. |
掲載区分 | 国内 |
出版社・発行元 | Japanese Journal of Applied Physics |
著者・共著者 | Kouhei Tanaka,Tomoji Yoshida, Kazuishi Sato, Toshio Kondo, Kazuhiro Yasukawa |
概要 | We present a blood-mimicking fluid for the Doppler test object of medical diagnostic instruments. Accurate measurement in a flow Doppler test requires a blood-mimicking fluid (BMF) that has the acoustic velocity and density defined in the International Electrotechnical Commission (IEC) standard, and furthermore, they must be stable over time. To formulate a fluid with the desired density and acoustic velocity, we have developed a new fluid made of glycerine and water soluble silicone oil. The BMF is dispersed polystyrene particles as scatters in the new fluid. The density of the liquid can be adjusted to maintain the density of the liquid at the same value as that of the polystyrene particles, thus ensuring neutral buoyancy of the particles. The blood-mimicking fluid was stable over a period of 2 weeks, during which the density and acoustic velocity did not change. |